Skip to main content

Table 1 Clinical data of the 10 patents implanted who received the Argus II system

From: Argus II retinal prosthesis for retinitis pigmentosa in the Middle East: The 2015 Pan-American Association of Ophthalmology Gradle Lecture

Patient no

Eye

Gender

Age (years)

Axial length (mm)

Lens Status

Peeling of posterior hyaloid

Postoperative complications

No. of functioning electrodes

No. of rehab sessions

1

Right

Male

45

24.96

Pseudophakic

YES

Conjunctival erosion

57

8

2

Left

Female

52

24.70

Aphakic

NO

Electrical Array Malposition

47 + 

7

3

Right

Female

29

22.76

Pseudophakic

NO

NO

59

6

4

Left

Male

45

25.83

Pseudophakic

YES

NO

56

3

5

Right

Male

38

23.75

Pseudophakic

NO

NO

37*

3

6

Left

Male

55

24.05

Pseudophakic

YES

Mild vitreous hemorrhage

55†

2

7

Right

Male

33

24.20

Phakic

YES

NO

58

5

8

Left

Male

47

22.76

Phakic

YES

High IOP

60

2

9

Right

Female

26

23.74

Pseudophakic

NO

NO

60

0

10

Right

Male

43

22.90

Phakic

NO

NO

59

0

  1.  + loss of 10 electrodes channels during electronic array tacking
  2. *18 electrodes with high impedance value were observed at the end of the surgery. Electrode channels were damaged during the extra ocular placement of the device
  3. After the tack insertion into the vitreous chamber, the surgeon observed that the tack was misaligned with the tack tool tip. When the surgeon removed it from the vitreous chamber, the tack dislodged from the tack tool and fell inside the vitreous cavity. The tack was managed as a foreign body and removal procedure was performed
  4. IOP Intraocular pressure, No. Number, mm millimeters, RE Right eye, LE Left eye, Gen Gender