New approaches to explanting high-viscosity silicone oil in retinal surgery—polyimide cannulas vs extraction sleeves vs a luer-trocar

Abstract

Background

The removal of high-viscosity silicone oil in retinal surgery is time-intensive. In this laboratory and porcine eye study, we evaluated the efficiency of novel extraction techniques, namely shortened polyimide-cannulas, an extraction sleeve and an experimental luer-trocar system, for commercially available silicone oils.

Methods

The volume of silicone oil after a 5-min removal time was quantified. The feasibility of all studied techniques was additionally tested in vitrectomized porcine eyes. Siluron 1000, 2000, 5000 and Xtra, as well as Densiron 68 and Densiron Xtra were examined.

Results

Shortening the 23G-polyimide-cannula from 6 mm up to 1.5 mm led to a significantly higher mean aspirated volume (e.g., 1 ml/min to 1.7 ml/min for Siluron Xtra). The extraction sleeve and luer-trocar increased the flow threefold compared to the shortest polyimide cannulas at 23G (5.2 ml/min and 5.2 ml/min vs 1.7 ml/min for Siluron Xtra, respectively). The extraction sleeve and luer-trocar system reached a similar flow at 27G as the best performing 23G-cannula. Angulation and movement of the eye was possible with the 2 mm-cannula, no angulation was possible when using the extraction sleeve. Stability and control were not impaired with the experimental luer-trocar setup.

Conclusion

Shorter polyimide-cannulas and extraction sleeves increase the extraction flow but may decrease stability during surgery. The luer-trocar system facilitates rapid removal while offering great stability and control. Small 27G silicone oil removal is possible and time efficient.

Background

Silicone oils are used as intraocular endotamponades separately by Amaly and Cibis in 1962 [1, 2] and have become a widespread and accepted long-term option for replacing the vitreous body after vitrectomy for retinal detachment, extensive diabetic retinopathy or endophthalmitis. While in human studies similar anatomic and visual outcomes were achieved with low- and high-viscosity silicone oils [3], high-viscosity silicone oils with viscosities of 5000mPas and above showed lower tendencies to emulsify in in-vitro studies [4,5,6,7,8]. Thus, many vitreoretinal surgeons tend to use high-viscosity silicone oils in most cases.

During surgery, a volume of 4–8 ml of oil is usually injected into and removed from the vitreous cavity after 3 to 12 months. Even at injection pressures of 6 bar this leads to injection times of more than 15 min in case of oils with added high molecular weight components [9, 10]. During silicone oil removal most commonly lower pressures between 600 and 700 mmHg are used further aggravating the surgical downside of these oils. Additionally, considering the upfront trend towards smaller gauge-surgery, new solutions to effectively remove silicone oil from the vitreous cavity are needed [11, 12]. Further, time-efficient techniques may not only translate to time savings during surgery but could also lighten the financial burden on hospitals.

The laminar flow through a tube can be estimated using the Hagen-Poiseuille equation. The following parameters determine the volumetric flow rate: The pressure difference between the two ends of the tube, the dynamic viscosity of the fluid, the length of the tube and, most importantly, the inner radius/ diameter of the tube. While certain factors, like the viscosity and the pressure difference, have been previously optimized or cannot be optimized, two factors may be improved during surgery, namely the length of the cannula as well as, most importantly, the inner diameter of the cannula. Previously, our group introduced polyimide-cannulas to the scientific literature on vitreoretinal surgery, as they can reduce injection times down to a quarter compared to other commercially available metal cannulas [10]. Polyimide is a rigid, high-performance polymer. The rigidness of the material allows for a slimmer wall thickness, increasing the inner diameter of the cannula while maintaining the same outer diameter. Figure 1 shows the removal techniques applied in this study.

Fig. 1
figure 1

Applied silicone oil removal techniques. a: Polyimide cannulas have a thinner wall due to the rigid high-performance polymer polyimide thus increasing the inner diameter. We tested 5 different lengths of the otherwise unchanged 23G-polyimide-cannula in a laboratory study and a porcine eye study. b: The extraction sleeve is placed over the trocar, thus using the inner lumen of the trocar for the silicone oil removal process. The use with 23G-, 25G- and 27G-trocars (schematically shown in blue, yellow and red, respectively) was evaluated. c: The luer-trocar is a system consisting of a female luer-lock piece attached to a trocar. As such, a stable connection to the syringe is formed. The trocar is used as the lumen. 23G-,25G- and 27G-luer trocar systems (schematically shown in blue, yellow and red, respectively) were used

We sought to evaluate if shortening the polyimide-cannula leads to a clinically relevant decrease of removal time. Additionally, another extraction technique are extraction sleeves that are placed over the trocar, thus effectively using the trocar as the lumen. Finally, we tested a prototype of a luer-trocar system: A female luer-lock part is directly attached to the trocar, allowing the direct connection of a syringe to the trocar, thus combining the benefits of cannulas (high control) with using the trocar as the removal lumen (high flow).

Materials and methods

Materials

Removal speed was evaluated for six commercially available silicone oils: Siluron 1000, Siluron 2000, Siluron Xtra, Siluron 5000, Densiron 68 and Densiron Xtra (all provided by Fluoron GmbH, Ulm, Germany). The compounds of the tested silicone oils are presented in Table 1. Throughout all experiments, 23G-, 25G- and 27G-trocars were used in conjunction with the removal technique to be tested.

Table 1 Compounds and density of examined silicone oils

Cannulas and extraction sleeves

The removal speed was evaluated for 23G-polyimide cannulas (provided by GEUDER AG, Heidelberg, Germany) in different lengths: 6 mm (standard length), 4 mm, 3 mm, 2 mm and 1.5 mm. Results were compared to the removal speed of an extraction sleeve (Alcon, Fort Worth, Texas, USA), which is placed over the trocar after removing the plastic valve. The sleeve was tested in combination with a 23G-, 25G- and a 27G-trocar (provided by GEUDER AG, Heidelberg, Germany). Finally, the prototype of a luer-trocar system was used. It consisted of a standard trocar (23G, 25G, 27G) connected to a female luer-lock part. Figure 1 depicts the different approaches to silicone oil removal tested in this study. A 10 ml syringe was connected to the extraction approaches mentioned above. The syringe was placed in a reusable silicone oil removal kit (Geuder AG, Heidelberg, Germany) connected to a Geuder Megatron S3 (Geuder AG, Heidelberg, Germany).

Methods

Experimental setup

A Megatron S3 (GEUDER AG, Heidelberg, Germany) was connected to a reusable silicone oil holder. Using the viscous fluid extraction mode, the vacuum was set to the device maximum of 600 mmHg. The vacuum pressure was constantly monitored throughout all experiments. A trocar was placed in a piece of porcine sclera to mimic the setting in the operating room. The respective cannula was inserted into the trocar, the extraction sleeve was placed on top of the trocar, or the luer-trocar system was placed directly in the sclera. Next, the vacuum was applied for 5 min. The total volume of silicone oil aspirated during these 5 min was quantified. Experiments were carried out threefold for every cannula length, for the extraction sleeve and the luer-trocar for each silicone oil, respectively. The experiments were carried out with the 5-min time limit to ensure that the maximum removal speed while the instrument tip is immersed in the silicone oil bubble is recorded. If 10 ml of an oil sample were aspirated in less than 5 min, the theoretical volume that could be aspirated in 5 min was extrapolated.

Porcine eye model

To assess if all removal techniques are clinically applicable, we conducted feasibility experiments in vitrectomized, silicone oil-filled pig eyes. The stability and the ability to use the extraction technique to angulate the trocar were evaluated. In this porcine eye model, silicone oil removal was fully completed. A second trocar with a balanced salt solution infusion line was placed and set to a pressure of 40 mmHg to ensure stable intraocular pressure.

Statistical analysis

To investigate the impact of the shortening of the polyimide-cannula, linear regression analyses were performed for each individual type of silicone oil. To determine the effect of the extraction sleeve and the experimental luer-trocar system on removal time, t-tests between the mean removal time of the shortest cannula and the setup to be tested were performed. Statistical analyses were performed using STATA17BE (StataCorp, College Station, Texas, USA) and Prism 8 (GraphPad Software Inc., San Diego, California, USA).

Results

The effect of the cannula length on removal speed

The volume aspirated in 5 min for all tested silicone oils and removal techniques is depicted in Fig. 2.

Fig. 2
figure 2

Mean extracted volume in 5 min using different techniques for a variety of clinically used silicone oils. Figure 2 depicts the volume aspirated in 5 min for all tested extraction techniques. Shortening the cannula from 6 mm to 1.5 mm increased the aspirated volume in a near-linear fashion. Extraction sleeves and the luer-trocar system showed significantly greater volumes aspirated in 5 min when compared to polyimide cannulas and even aspirated a comparable volume at 27G. No significant difference in removal speed was apparent between the extraction sleeve and the luer-trocar system

All commercially available silicone oils could be aspirated with all tested removal techniques. The monitored pressure of the applied vacuum reached its maximum of 600 mmHg roughly 10–15 s after the start of the removal process. For all silicone oils, the shortening of the cannula significantly increased the volume aspirated in 5 min in a near-linear fashion (r2 = 0.980, 0.994, 0.971, 0.946, 0.958 and 0.993 for Siluron 1000, Siluron 2000, Siluron Xtra, Siluron 5000, Densiron 68 and Densiron Xtra, respectively). Comparing the shortest to the longest 23G-polyimide cannula, a mean increase in removal speed of 1.57 ml/min, 0.90 ml/min, 0.7 ml/min, 0.63 ml/min, 0.87 ml/min and 2.73 ml/min was observed for Siluron 1000, Siluron 2000, Siluron Xtra, Siluron 5000, Densiron 68 and Densiron Xtra, respectively.

Removal speed of the extraction sleeve and the experimental luer-trocar setup

The use of the extraction sleeve and the luer-trocar system majorly increased the aspirated volume of silicone oil. Depending on the tested oil, the removed volume was up to fourfold increased (Siluron 5000) when using a 23G-trocar compared to the shortest (1.5 mm) 23G-cannula. The mean increase in removal speed was 13 ml/min for Siluron 1000, 9.1 ml/ min for Siluron 2000, 3.6 ml/min for Siluron Xtra, 3.1 ml/min for Siluron 5000, 9.1 ml/min for Densiron 68 and 13.8 ml/min for Densiron Xtra, respectively. (p < 0.001 for all oils, unpaired t-tests, see Fig. 2). At 25G, the extraction sleeve and the luer-trocar still outperformed the 23G-cannula significantly by aspirating twice the volume. (p < 0.001, unpaired t-test, see Fig. 2) Even at 27G, the luer-trocar and extraction sleeve performed similarly to the shorted polyimide cannula at 23G allowing the time efficient extraction of silicone oil at a small gauge. For all tested silicone oils, the removal speed of the luer-trocar system was not statistically different compared to the extraction sleeve. (p > 0.90 for all oils).

Feasibility experiments in vitrectomized porcine eyes

During silicone oil removal surgery, it may be necessary to angulate the trocar or the eye using the inserted cannula to aspirate the remaining silicone oil bubble. Thus, we evaluated all polyimide-cannulas (6 mm, 4 mm, 3 mm, 2 mm, 1,5 mm), the extraction sleeve and the luer-trocar system in a porcine eye feasibility study. Angulating the trocar was not impaired for cannula lengths of 3, 4 and 6 mm. While smaller movements without the dislocation of cannula tip were still possible with a length of 2 mm, a length of 1.5 mm proved to come along with reduced control. Similarly, the extraction sleeve did not allow the positioning of the trocar aperture. As silicone oil may act as a lubricant, caution must be taken when using the extraction sleeve to not lose contact with the trocar. The luer-trocar system was stable and allowed the angulation of the trocar with ease. Figure 3 depicts the application of the 23G-polyimide cannula (3 mm), the extraction sleeve and the luer-trocar system.

Fig. 3
figure 3

Feasibility experiments on porcine eyes a: Unproblematic insertion of a 3 mm 23G-polyimide cannula. The angulation of the trocar to place the trocar tip in the remaining oil bubble was easily achievable. b: Placement of the extraction sleeve over a 23G-trocar. In comparison to the shortest cannula, a fourfold increase of the volume aspirated in 5 ml was observed. Angulation of the trocar is not possible. Surgeons must be careful to not lose contact to the trocar during removal. c: The luer-trocar system allowed easy angulation of the trocar tip within the eye. The luer-trocar system had similar removal speeds as the extraction sleeve but offered preferable stability and surgical control

Discussion

In this study, we evaluated three techniques to remove silicone oil from the vitreous cavity. A steady increase of flow was observed when shortening the cannula from 6 mm to 1.5 mm, the use of an extraction sleeve or a luer-trocar system massively decreased the time to explant a variety of silicone oils. Although all solutions significantly decreased removal time, there may be some surgical drawbacks due to a decreased ability to angulate the corresponding trocar. This was apparent when using the 1.5 mm 23G-polyimide cannula as well as the extraction sleeve. The third technique, a luer-trocar system that allows a direct connection of the trocar with any medical syringe, combined the advantages of both techniques.

To this date, few laboratory research has been conducted to optimize surgical gear when using high-viscosity silicone oils. Williams et al. in [9] made an effort to improve the injectability and removal of silicone oil by using the effect of “shear thinning”. In fact, the present study includes 3 silicone oils, namely Siluron 2000, Siluron Xtra and Densiron Xtra, that make use of this approach. By creating a high-viscosity silicone oil by mixing low-viscosity silicone oil with high molecular weight components instead of using a one component silicone oil with a medium chain length, favorable effects of the injectability can be achieved while still showing comparable or even better results in the tendency to emulsify [5, 13]. In this study, we focused on improving the injection and removal of silicone oil from the vitreous cavity by shortening the improved cannula made from a rigid, high-performance polymer named polyimide [10]. Without any change to the standard operating procedures of a silicone oil removal surgery we saw an increase in flow of up to 158% (Siluron 5000) when using a shorter cannula. While already a substantial amount of time can be saved by using a polyimide-cannula instead of conventional metal cannulas [10], the shortening decreases the removal time in a scenario of 5 ml of Siluron 5000 by an additional 8 min.

Further, we tested the use of extraction sleeves. While some authors have suggested its use [12], to our knowledge, its performance have never been systematically compared to other silicone oil removal techniques in a controlled laboratory setting. While the increase in flow must be acknowledged, the extraction sleeve also holds three weaknesses that became apparent during the feasibility study. 1. It does not allow the angulation of the trocar tip. 2. The connection between the sleeve and the trocar is susceptible to disconnection through small movements. 3. The extraction sleeve can only be used during the removal process and not during silicone oil injection.

The third technique evaluated in this study is a prototype model of a luer-trocar. The luer-trocar offered optimal conditions for silicone oil removal and may offer surgical innovations not limited to silicone oil surgery in the future. A tight connection to a tube or syringe may improve other current solutions in vitreoretinal surgery. A possible application could be the infusion flow. Indeed, while a luer-trocar system is not commercially available, the recently developed EVA Aveta trocar system by DORC features a Push-Fit connection that shows similar traits to the luer-trocar system studied here and is currently used for fluid infusion during surgery. With small adjustments, specifically a connecting piece between the push-fit trocar and a luer-syringe, the system could enable a luer-trocar system for silicone oil removal to be quickly implemented in clinical practice.

In the past, multiple clinical studies dealt with the removal of silicone oils. Ohayon et al. [14] presented an inexpensive two-trocar setup. However, the oil type used is not mentioned in the article and the exact setup of the active removal trocar is not well described, as the focus of the study is the safety of a two-trocar approach. Priester et al. compared two-trocar approaches in a clinical study. They used a hybrid approach with one 23G-port and one 25G-port compared to a standard 23G-approach. Again, in the article, the used cannula or system for the active removal trocar is not mentioned. As such, again no real comparison of results is possible [15]. Another article on silicone removal was published by Patwardhan et al. in [16]. Goal of this study was to compare a passive, sutured 20G-approach with a passive 23G sutureless approach. As our study uses active removal and focuses on the usage of the trocar rather than their gauge itself, our study is able to expand on this knowledge. Similarly, Jusufbegovic in [17] showed that even a 25G sutureless approach is more time efficient compared to a 20G approach requiring sutures. Again, the optimal usage of the extraction trocar was not the main study question. Song et al. [18] introduced an early version of an extraction sleeve. For that, they modified a blood transfusion system while using a 23G- setup with 3 trocars. Similarly, to the extraction sleeve in this study, the valve has to be removed. However, this approach was not compared to other techniques. Zhang et al. in [19] presented a similar extraction sleeve technique with a 27G setup. However, they did not use an automatically controllable approach to generate vacuum but used a 10 ml syringe in which the plunger is lifted by surgeon to generate vacuum. The approach requires both hands of the surgeon to be involved and vacuum generated by this approach was not quantified appropriately. Finally, the only systematic approach comparing a cannula and an extraction sleeve under controlled conditions was presented by Tabandeh et al. at ARVO 2014. However, the group focused on low-viscosity silicone oil with 1000 mPas. The only study presenting a similar approach to our experimental luer-trocar was published by Akkan et al. [20]. The authors present a common 19G-extraction cannula with a thickened tip that allows the occlusion of a 23G-trocar. In the luer-trocar presented here, a direct connection between the luer-setup and the trocar is provided, the 19G-extraction cannula is interposed in the setup by Akkan et al. [20]. Both approaches provide benefits and should be further evaluated in a comparison study.

In the meanwhile, possible hybrid approaches that first use an extraction sleeve followed by an approach with a cannula to remove the residual silicone oil may be a good option in clinical practice for high-viscosity silicone oils. Due to the greater time savings when explanting high-viscosity silicone oils, possible delays caused by exchanging the sleeve for the cannula may be tolerable.

In conclusion, high-viscosity silicone oil removal can be expedited using shortened polyimide-cannulas, extraction sleeves as well as luer-trocars in the near future. These time savings may translate to shorter surgery times for patients and are easy to implement in daily clinical practice.

Availability of data and materials

Data is available from the corresponding author upon reasonable request.

References

  1. Armaly MF. Ocular tolerance to silicones. I. Replacement of aqueous and vitreous by silicone fluids. Arch Ophthalmol. 1962;68:390–5.

    Article  CAS  PubMed  Google Scholar 

  2. Cibis PA, Becker B, Okun E, Canaan S. The use of liquid silicone in retinal detachment surgery. Arch Ophthalmol. 1962;68:590–9.

    Article  CAS  PubMed  Google Scholar 

  3. Scott IU. Outcomes of complex retinal detachment repair using 1000- vs 5000-centistoke silicone oil. Arch Ophthalmol. 2005;123:473.

    Article  CAS  PubMed  Google Scholar 

  4. Caramoy A, Schröder S, Fauser S, Kirchhof B. In vitro emulsification assessment of new silicone oils. Br J Ophthalmol. 2010;94:509–12.

    Article  PubMed  Google Scholar 

  5. Chan YK, Czanner G, Shum HC, Williams RL, Cheung N, Wong D. Towards better characterization and quantification of emulsification of silicone oil in vitro. Acta Ophthalmol. 2017;95:e385–92.

    Article  CAS  PubMed  Google Scholar 

  6. Crisp A, de Juan E, Tiedeman J. Effect of silicone oil viscosity on emulsification. Arch Ophthalmol. 1987;105:546–50.

    Article  CAS  PubMed  Google Scholar 

  7. Heidenkummer HP, Kampik A, Thierfelder S. Emulsification of silicone oils with specific physicochemical characteristics. Graefes Arch Clin Exp Ophthalmol. 1991;229:88–94.

    Article  CAS  PubMed  Google Scholar 

  8. Hammer M, Schickhardt S, Munro DJ, Scheuerle A, Mayer CS, Auffarth GU. Physicochemical properties of explanted silicone oil after use as an intraocular tamponade. Trans Vis Sci Tech. 2022;11:3.

    Article  Google Scholar 

  9. Williams RL, Day MJ, Garvey MJ, Morphis G, Irigoyen C, Wong D, et al. Injectability of silicone oil-based tamponade agents. Br J Ophthalmol. 2011;95:273–6.

    Article  CAS  PubMed  Google Scholar 

  10. Hammer M, Schickhardt SK, Munro D, Khoramnia R, Scheuerle A, Mayer CS, et al. The novel use of high-flow polyimide cannulas to improve silicone oil injectability in vitreoretinal surgery. Retina. 2022. https://doi.org/10.1097/IAE.0000000000003432.

    Article  PubMed  Google Scholar 

  11. Riemann CD, Miller DM, Foster RE, Petersen MR. Outcomes of transconjunctival sutureless 25-gauge vitrectomy with silicone oil infusion. Retina. 2007;27:296–303.

    Article  PubMed  Google Scholar 

  12. Tabandeh H, Boscia F, Boyer DS. Removal of silicone oil with 25 gauge and 23 gauge cannula systems. Invest Ophthalmol Vis Sci. 2014;55:2342–2342.

    Google Scholar 

  13. Williams RL, Day M, Garvey MJ, English R, Wong D. Increasing the extensional viscosity of silicone oil reduces the tendency for emulsification. Retina. 2010;30:300–4.

    Article  PubMed  Google Scholar 

  14. Ohayon A, Esa S, Rubowitz A. A two-port inexpensive and effective method for silicone oil removal. Isr Med Assoc J. 2020;22:89–93.

    PubMed  Google Scholar 

  15. Priester LD, Laspas P, Stoffelns B, Pfeiffer N. Removal of higher viscosity oil through a hybrid 2-port 23/25 gauge pars plana vitrectomy. Ophthalmologe. 2021;118:394–6.

    Article  CAS  PubMed  Google Scholar 

  16. Patwardhan SD, Azad R, Shah V, Sharma Y. The safety and efficacy of passive removal of silicone oil with 23-gauge transconjunctival sutureless system. Retina. 2010;30:1237–41.

    Article  PubMed  Google Scholar 

  17. Jusufbegovic D, Gambrell JD, Barak Y, Schaal S. Twenty-five-gauge vitrectomy for the removal of 5000 centistokes silicone oil. Retina. 2013;33:1938–42.

    Article  CAS  PubMed  Google Scholar 

  18. Song Z-M, Hu X-T, Wang L, Hu Z-X, Zhao P-Q, Chen D. A modified approach to actively remove high viscosity silicone oil through 23-gauge cannula. Int J Ophthalmol. 2016;9:1294–8.

    PubMed  PubMed Central  Google Scholar 

  19. Zhang Z, Zhang S. Novel and easy techniques for 27-gauge silicone oil infusion and removal. Int Ophthalmol. 2019;39:1543–51.

    Article  PubMed  Google Scholar 

  20. Akkan F, Dinc D. A novel design silicone oil removal cannula. Asian J Med Sci. 2021;12:39–43.

    Article  Google Scholar 

Download references

Acknowledgements

We would like to thank the GEUDER AG, Heidelberg, Germany for providing us with cannulas. We would like to thank Fluoron GmbH, Ulm, Germany for supplying us with silicone oils. We would like to thank Lee Embray for his help creating luer-trocar prototypes.

None of the other authors declares a possible conflict of interest.

Funding

For the publication fee we acknowledge financial support by Deutsche Forschungsgemeinschaft within the funding programme “Open Access Publikationskosten” as well as by Heidelberg University. This study was made possible through the support of the Klaus Tschira Foundation granted to GUA. The funding organization had no role in the design or conduct of this research. This study was supported by the Ernst und Berta Grimmke-Foundation granted to MH (Grant No. 14/22). Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

MH, SS, PU, GUA designed the study. MH, DM and GUA conducted the analysis. MH, SS, DM, AS, RK, PU and GUA were involved in writing of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Gerd U. Auffarth.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

Gerd Uwe Auffarth reports grants, personal fees, non-financial support and consulting fees from Johnson & Johnson and Alcon, grants, personal fees, and non-financial support from Carl Zeiss Meditec, Hoya, Kowa,Oculentis/Teleon, Rayner, Santen, Sifi, Ursa-pharm, grants and personal fees from Biotech, Oculus, EyeYon grants from Acufocus, Conta-mac, Glaukos, Physiol, Rheacell, outside the submitted work. Ramin Khoramnia reports grants, personal fees, and non-financial support from Alcon, Johnson&Johnson, Hoya, Physiol, Rayner, personal fees, and non-financial support from Kowa, Ophtec, Oculentis/Teleon, Santen, and Acufocus, outside the submitted work.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hammer, M., Schickhardt, S., Munro, D. et al. New approaches to explanting high-viscosity silicone oil in retinal surgery—polyimide cannulas vs extraction sleeves vs a luer-trocar. Int J Retin Vitr 9, 43 (2023). https://doi.org/10.1186/s40942-023-00481-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40942-023-00481-w

Keywords