Image sharpening algorithms improve clarity of surgical field during 3D heads-up surgery

Abstract

Background

Image-sharpening algorithms with color adjustments enable real-time processing of the surgical field with a delay of 4 msec for heads-up surgery using digital three-dimensional displays. The aim of this study was to investigate the usefulness of the algorithms with the Artevo 800® digital microscope.

Methods

Seven vitreoretinal surgeons evaluated the effects of image-sharpening processing on the clarity of the surgical field with the Artevo 800® system that is used for cataract and vitreous surgeries. The scorings were made on a 10-point scale for anterior capsulotomy, phacoemulsification, cortex aspiration, core vitrectomy, and peeling of an epiretinal membrane or an internal limiting membrane. In addition, the images during the internal limiting membrane peeling were processed with or without color adjustments. We also evaluated the skewness (asymmetry in the distribution of the pixels) and kurtosis (sharpness in the distribution of the pixel) of the images to evaluate the contrast with each intensity of image-sharpening.

Results

Our results showed that the mean visibility score increased significantly from 4.9 ± 0.5 at 0% (original image) to 6.6 ± 0.5 at 25% intensity of the image-sharpening algorithm (P < 0.01). The visibility scores of the internal limiting membrane increased significantly from 0% (6.8 ± 0.3, no color adjustments) to 50% after the color adjustments (7.4 ± 0.4, P = 0.012). The mean skewness decreased significantly from 0.83 ± 2.02 at 0% (original source) to 0.55 ± 1.36 at 25% intensity of the image-sharpening algorithm (P = 0.01). The mean kurtosis decreased significantly from 0.93 ± 2.14 at 0% (original image) to 0.60 ± 1.44 at 25% intensity of the image-sharpening algorithm (P = 0.02).

Conclusions

We conclude that the image-sharpening algorithms can improve the clarity of the surgical field during 3D heads-up surgery by decreasing the skewness and kurtosis.

Trial registration

This was a prospective clinical study performed at a single academic institution, and the procedures used were approved by the Institutional Review Committee of the Kyorin University School of Medicine (reference number, 1904). The procedures also conformed to the tenets of the Declaration of Helsinki.

Background

Heads-up surgery (HUS) using digital three-dimensional displays has been introduced for ophthalmic surgeries [1,2,3,4,5,6,7,8,9,10,11,12,13]. In HUS, surgeons perform surgical procedures while looking not into the eyepieces of conventional surgical microscopes but at three-dimensional displays placed in front of them displaying images of the surgical field from a high-resolution dual-camera system. Many studies have reported that HUS is a very good option for cataract and vitreoretinal surgeries [1,2,3,4,5,6,7,8,9,10,11]. Eckardt [5] reported that HUS is well suited for vitreoretinal procedures in terms of superior ergonomics and enhancement of the brightness of the surgical field without exposing the retina to excessive light levels. The improved sensitivity of video cameras has enabled surgeons to perform surgeries under low-light conditions. This is another advantage of HUS for vitreous surgery which reduces retinal phototoxicity [14,15,16,17]. A comparison of 3D HUS to conventional surgery using a surgical microscope revealed that the outcomes of HUS are equal to or better than those of conventional surgery [3, 4, 8, 13, 17].

Image-sharpening technology is applied in surveillance cameras and is designed to sharpen images in backlighted and nighttime conditions. Hoshi and associates [18] reported that image processing with comb-removal and image-sharpening algorithms improved the visibility of the surgical field during dacryoendoscopy. Tasaki and associates [19] reported that image processing with comb-removal and image-sharpening algorithms improved the visibility of the surgical field significantly during 27-gauge endoscopic vitrectomy. We investigated the usefulness of image-sharpening algorithms during HUS with the Ngenuity® 3D Visualization System [20].

The purpose of this study was to determine whether image-sharpening algorithms can improve the clarity of the surgical field during HUS with the Artevo 800® digital microscope which does not have High Dynamic Range (HDR) video cameras. We also examined whether a color adjustment algorithm can further improve the clarity of the surgical field during HUS.

Methods

Subjects

3D HUS was performed during 5 vitreoretinal and 6 cataract surgeries on 6 eyes of 6 patients by a single surgeon (MI). The surgeries were performed with the Artevo 800® (Carl Zeiss Meditec, Oberkochen, Germany) and the medical image enhancer (MIEr®, Logic & Design, Tokyo, Japan) at 25% intensity of the image-sharpening algorithm with or without 50% color adjustments.

Surgical procedures

The Constellation® Vision System (Alcon Laboratories, Fort Worth, TX) was used for the 27-gauge pars plana vitrectomy (PPV) and cataract surgery. The posterior hyaloid cortex was made more visible by an intravitreal injection of triamcinolone acetonide (MaQaid®, Wakamoto Pharmaceutical Co., LTD, Tokyo, Japan). Brilliant blue green (BBG) was used to enhance the visibility of the internal limiting membrane (ILM).

A video of the surgery was recorded with the HVO-3300MT® (Sony Group Corporation, Tokyo, Japan) without any image-sharpening. We analyzed the images from the intraoperative videos of the different surgical procedures: anterior capsulotomy with continuous curvilinear capsulorhexis (CCC), lens removal with phacoemulsification and aspiration (PEA), cortex removal with irrigation and aspiration (IA), core vitrectomy (Vit), removal of an epiretinal membrane (ERM), and peeling of the ILM stained with BBG.

To evaluate the effect of the image-sharpening algorithms, the original video images were processed with the medical image enhancer at 25% intensity of the image-sharpening and compared to the original images. The processed values of the image-sharpening intensity ranged from 0 to 255 and the setting of 25% intensity was 64/255 when expressed as a ratio. We analyzed 6 images each during CCC, PEA, and IA, 8 images of core vitrectomy, ERM peeling, and ILM peeling. To evaluate the effect of color adjustments, 9 images of ILM peeling from the same videos after an intravitreal injection of BBG were processed with the medical image enhancer at 50% color adjustments with 25% image-sharpening intensity and compared with no color adjustments at 25% intensity of image-sharpening. The color rate value of the color adjustments ranged from 0 to 64 and the color rate of 16 represented 1.0 (0% color adjustments) and 32 represented 1.5 times (50% color adjustments).

Visibility scores

Seven vitreoretinal surgeons of intermediate or higher surgical skills (KN, MI, AT, YY, MM, TK, and TI) examined and scored these 43 still images on a scale of 0 to 10 points with 0 as the poorest and 10 as the clearest depending on how clear scorers could see the objects they were manipulating, and how well the objects were contrasted against the background after the information of the patients was masked. The visibility scores were evaluated between the still images with and without 25% intensity of the image-sharpening algorithms. The visibility scores were also evaluated between the still image with and without 50% color adjustments at 25% intensity of the image-sharpening algorithms.

Skewness and kurtosis

The skewness and kurtosis of those images were calculated using the ImageJ Fiji software [21] to evaluate how the image-sharpening algorithms sharpened the images. The skewness is a measure of asymmetry, i.e., how the distribution of the pixels is pulled to one or the other side of the peak intensity [22]. For instance, if the skewness is negative, the histogram is negatively skewed, and the frequency over the darker intensities is spread wider. When the absolute value of the skewness is near zero, the image is balanced and not extremely dark or bright. Thus, if the skewness of the surgical images is far from zero, the surgeon can perceive the images as being too dark or too bright to do the surgical procedures. The kurtosis of an image is how high and sharp the central peak of the histogram of the brightness of the pixels is [22]. If the kurtosis is high, there are many pixels that have the same pixel value in the image. Thus, the image is of low contrast.

Statistical analyses

The Wilcoxon signed-rank test was used to compare the two groups. The intra-rater reliability among the examiners was evaluated with the intraclass correlation coefficient (ICC). All statistical analyses were performed using SPSS (version 28.0; IBM, Armonk, New York, NY, USA).

Results

All surgeries were performed successfully without intraoperative complications with the Artevo 800® and the medical image enhancer with the image-sharpening algorithms at 25% intensity or 25% intensity of image-sharpening with 50% color adjustments during ILM peeling with BBG staining.

Visibility scores

The mean visibility score increased significantly from 4.9 ± 0.5 at 0% for the original image to 6.6 ± 0.5 at 25% intensity for the enhanced image (P < 0.01, Wilcoxon signed rank test, Fig. 1). The visibility scores for CCC, PEA, IA, Vit, and ERM increased significantly from 0 to 25% intensity of the image-sharpening (CCC, P = 0.027; PEA, P = 0.027; IA, P = 0.028; Vit, P < 0.01; ERM, P = 0.012; ILM, P < 0.01). The visibility score for the ILM with color adjustments increased significantly from 0% (6.8 ± 0.3, no color adjustments) to 50% after the color adjustments (7.4 ± 0.4, P = 0.012). The ICC of the visibility scores without the image-sharpening algorithms and with 25% intensity was 0.760 (95% CI: 0.506; 0.943, P < 0.001) and 0.802 (95% CI: 0.564–0.955, P < 0.001) during CCC, 0.851 (95% CI: 0.650; 0.967, P < 0.001), and 0.831 (95% CI: 0.622; 0.962, P < 0.001) during PEA, 0.830 (95% CI: 0.606; 0.962, P < 0.001), and 0.850 (95% CI: 0.639; 0.967, P < 0.001) during IA, 0.765 (95% CI: 0.556; 0.942, P < 0.001), and 0.720 (95% CI: 0.496; 0.928, P < 0.001) during Vit, 0.726 (95% CI: 0.470; 0.932, P < 0.001), and 0.580 (95% CI: 0.300; 0.880, P < 0.001) during ERM, 0.868 (95% CI: 0.719; 0.970, P < 0.001), and 0.781 (95% CI: 0.578; 0.947, P < 0.001) during ILM, respectively. The ICC of the visibility scores without the color adjustments and with 50% color adjustments was 0.692 (95% CI: 0.462; 0.918, P < 0.001), and 0.693 (95% CI: 0.463; 0.918, P < 0.001), respectively. The intra-rater reliability among the examiners was between moderate agreement (0.41–0.60), substantial agreement (0.61–0.80), and almost perfect agreement (0.81–1.00).

Fig. 1
figure 1

Mean visibility scores for cataract and vitreous surgeries. The mean visibility score of all procedures increases significantly with 25% intensity of the image-sharpening. The visibility scores for CCC, PEA, IA, Vit, and ERM increase significantly with 25% intensity. CCC = continuous circular capsulorhexis, PEA = phacoemulsification and aspiration, IA = irrigation and aspiration, Vit = core vitrectomy, ERM = epiretinal membrane peeling, ILM = internal limiting membrane peeling

Image clarity during surgeries

For CCC, the edge of the anterior capsule was easier to detect after increasing the intensity of the image sharpening (Fig. 2). For PEA, the edges of the lens nucleus and the remaining cortex were more highlighted with the increase of intensity of image sharpening. For IA, the edge of the anterior capsulotomy and the remaining cortex, and wrinkles of the posterior capsule were seen clearer with increasing intensity of the image-sharpening. For vitrectomy, the area of the vitreous cavity illuminated by a light pipe was larger, and the retinal and choroidal vessels appeared more clearly after increasing the intensity of the image sharpening program (Fig. 3). For ERM peeling, the folds of the ERM appeared clearer with image sharpening, and for ILM peeling, the edges of ILM that were stained blue with BBG appeared more clearly with the image sharpening. For ILM peeling with color adjustments, the peeled ILM appeared more clearly with the color adjustments that was observed the yellow color was emphasized and the color tone was observed clearer with the color adjustments.

Fig. 2
figure 2

Surgical images of cataract surgery with and without image-sharpening algorithms. The edge of anterior capsulotomy is easily identified during continuous circular capsulorhexis (CCC) with 25% intensity of the image-sharpening. For phacoemulsification and aspiration (PEA) and irrigation and aspiration (IA), the edges of the lens nucleus and the remaining cortex are better highlighted with 25% intensity

Fig. 3
figure 3

Surgical images of vitreous surgery with and without image-sharpening algorithms. The lightened area of the vitreous cavity is larger and the retinal and choroidal vessels appear more clearly with 25% intensity during vitrectomy. The folds of the epiretinal membrane (ERM) and internal limiting membrane (ILM) stained blue with brilliant blue G appear more clearly with 25% intensity during ERM and ILM peeling. The peeled ILM appears more clearly with the color adjustments that enhance the yellow color of the ILM stained blue with brilliant blue G

Skewness and kurtosis

To evaluate the contrast of the images, we calculated the skewness and kurtosis of the same images. The mean skewness decreased significantly from 0.83 ± 2.02 at 0% for the original images to 0.55 ± 1.36 at 25% intensity of the image after the image-sharpening algorithm (P = 0.01, Table 1). The skewness for PEA, IA, Vit, and ERM decreased significantly when the intensity was changed from 0 to 25% of the image-sharpening algorithm (PEA, P = 0.027; IA, P = 0.028; Vit, P < 0.001; ERM, P = 0.012) but the change in the skewness was not significant for CCC, ILM, and ILM with color adjustments.

Table 1 Skewness of surgical images

The mean kurtosis decreased significantly from 0.93 ± 2.14 at 0% for the original image to 0.60 ± 1.44 at 25% intensity of the image-sharpening algorithm (P = 0.02, Table 2). The kurtosis for CCC, PEA, Vit, ERM, and ILM decreased significantly when the intensity was increased from 0 to 25% of the image-sharpening algorithm (CCC, P = 0.028; PEA, P = 0.027; Vit, P < 0.001; ERM, P = 0.012; ILM, P = 0.014). However, it was not significant for IA and ILM with color adjustments.

Table 2 Kurtosis of surgical images

Discussion

The results showed that the medical image enhancer improved the clarity of the surgical field for CCC, PEA, IA, vitrectomy, ERM peeling, and ILM peeling. In addition, the quantification of the distribution of the pixel intensities for these surgical images showed that the medical image enhancer improved the surgical clarity during cataract and vitrectomy surgeries. The image-sharpening algorithms were tuned to improve the clarity of live images by increasing the dynamic range per pixel that was calculated in real-time by 0.004 Sects. [18, 19]. The scoring by the surgeons showed that the image-sharpening algorithms improved the clarity of the surgical field by increasing the contrast between the object and the background. During cataract surgery, the image-sharpening algorithms showed the edges of the anterior capsulotomy, the wrinkles of the posterior capsule, and the edges of the residual lens cortex more clearly. For vitreoretinal surgery, the image-sharpening algorithms highlighted the vitreous, folds of the ERM, and the border of the areas with ILM that was not peeled and without ILM after it was peeled.

A histogram of the pixel distribution is an important tool for image processing and it gives a graphical representation of the distribution of pixel intensities in a digital image [22]. The skewness and kurtosis are used to evaluate the medical images such as in computed tomography (CT) scans, breast thermographs, and magnetic resonance imaging (MRI) [23,24,25]. We measured the skewness and kurtosis of the images and compared them with or without the image-sharpening algorithms. The image-sharpening algorithms lowered the skewness into the positive range and brought them closer to zero. This means that the algorithms shifted the bins toward the peak of the histogram [22]. The skewness of surgical images showed that the effectiveness of 25% image sharpening was not too high for procedures using a uniform bright microscope light source. However, for vitrectomy (light guide light source), the enhancement was very effective due to the characteristics of the light from a point source. Also, the image-sharpening algorithms lowered the kurtosis. This means that the algorithms widened and flattened the distribution of the bins of the histogram. The results of the measurements can be interpreted as image-sharpening algorithms that improved the clarity of the surgical images.

The gain and gamma corrections can adjust the overall brightness and darkness uniformly, but they cannot make adjustments when the bright and dark areas are mixed in the same image [26]. Artevo 800® has two integrated 4 K video cameras which are not High Dynamic Range (HDR) video cameras, so the video cameras cannot improve the dynamic range [27, 28]. If the brightness is calculated and set by referring to the histogram of luminance values in a small surrounding range for each pixel unit, the dynamic range can be optimized up to all corners of the screen. In this way, the dark areas are not “blacked out” and bright areas are not “whited out”. The image-sharpening algorithms can be designed to perform this process at high speed and to suppress noise generation and contrast reduction, making it possible to sharpen the images for practical use.

The Ngenuity® 3D Visualization System has dual HDR 2 K video cameras and a color filter system which enables surgeons to adjust the color tone of the surgical image. In contrast, Artevo 800® does not have HDR capability and a color filter system. For these reasons, we experimented with the image-sharpening algorithms for 3D HUS with the Artevo 800®. The results showed that the image-sharpening algorithms significantly improved the visual scores for various surgical procedures in simultaneous cataract and vitreous surgery even without HDR capability. The medical image enhancer improved the surgical experience during ILM peeling using the color adjustments by the medical image enhancer even without the color filter function.

This study has limitations. First, we did not assess the quality of the images processed using the medical image enhancer to compare them to the images without the image enhancer during a real-time surgery. Second, we did not assess the visual outcomes on whether the medical image enhancer improved the visual outcomes because we had only a few cases. Further studies are needed to determine whether the proposed algorithm contributes to improved surgical outcomes.

Conclusions

The image sharpening algorithms with a real-time processing of live images improves the intraoperative clarity during 3D HUS by decreasing skewness and kurtosis. The color adjustments by the medical image enhancer improve the surgical clarity during ILM peeling with BBG staining even without capability of color filter system.

Data Availability

The data presented in this study are available on request from the first author (K.N.).

Abbreviations

HUS:

Heads-up surgery

HDR:

high dynamic range

PPV:

pars plana vitrectomy

BBG:

brilliant blue green

ILM:

internal limiting membrane

CCC:

continuous circular capsulorhexis

PEA:

phacoemulsification and aspiration

IA:

irrigation and aspiration

ERM:

epiretinal membrane

CT:

computed tomography

MRI:

magnetic resonance imaging

References

  1. Agranat JS, Miller JB, Douglas VP, Douglas KAA, Marmalidou A, Cunningham MA, et al. The scope of Three-Dimensional Digital visualization Systems in vitreoretinal surgery. Clin Ophthalmol. 2019;13:2093–26. https://doi.org/10.2147/OPTH.S213834.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Ta Kim D, Chow D. The effect of latency on surgical performance and usability in a three-dimensional heads-up display visualization system for vitreoretinal surgery. Graefes Arc Clin Exp Ophthalmol. 2022;260:471–6. https://doi.org/10.1007/s00417-021-05388-6.

    Article  Google Scholar 

  3. Palácios RM, de Carvalho ACM, Maia M, Caiado RR, Camilo DAG, Farah ME. An experimental and clinical study on the initial experiences of brazilian vitreoretinal surgeons with heads-up surgery. Graefes Arch Clin Exp Ophthalmol. 2019;257:473–83. https://doi.org/10.1007/s00417-019-04246-w.

    Article  PubMed  Google Scholar 

  4. Del Turco C, D’Amico Ricci G, Dal Vecchio M, Bogetto C, Panico E, Giobbio DC et al. Heads-up 3D eye surgery: Safety outcomes and technological review after 2 years of day-to-day use. Eur J Ophthalmol. 2021, Apr30:11206721211012856. doi: https://doi.org/10.1177/11206721211012856. Epub ahead of print.

  5. Eckardt C, Paulo EB, HEADS-UP SURGERY FOR VITREORETINAL. PROCEDURES: an experimental and clinical study. Retina. 2016;36(1):137–47. https://doi.org/10.1097/IAE.0000000000000689.

    Article  PubMed  Google Scholar 

  6. Freeman WR, Chen KC, Ho J, Chao DL, Ferreyra HA, DEPTH OF FIELD, AND PHYSICIAN SATISFACTION DURING DIGITALLY ASSISTED VITREORETINAL SURGERY. Retina. 2019;39(9):1768–71. https://doi.org/10.1097/IAE.0000000000002236.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Brooks CC, Kitchens J, Stone TW, Riemann CD. Consolidation of Imaging Modalities Utilizing Digitally Assisted Visualization Systems: The Development of a Surgical Information Handling Cockpit. Clin Ophthalmol. 2020 Feb27;14:557–69. doi: https://doi.org/10.2147/OPTH.S239339.

  8. Coppola M, La Spina C, Rabiolo A, Querques G, Bandello F. Heads-up 3D vision system for retinal detachment surgery.Int J Retina Vitreous. 2017 Nov20;3:46. doi: https://doi.org/10.1186/s40942-017-0099-2.

  9. Babu N, Kohli P, Jena S, Ramasamy K. Utility of digitally assisted vitreoretinal surgery systems (DAVS) for high-volume vitreoretinal surgery centre: a pilot study. Br J Ophthalmol. 2020;104(3):432–6. https://doi.org/10.1136/bjophthalmol-2019-314123.

    Article  PubMed  Google Scholar 

  10. Ramesh PV, Balasubramaniam P, Devadas AK, Ray P, Ramesh SV, Ramesh MK, et al. Venturing into the third dimension of cataract surgery/observations with digital heads-up display - A personal take on NGENUITY for anterior segment surgeries. Indian J Ophthalmol. 2022;70(2):694–5. https://doi.org/10.4103/ijo.IJO_3008_21.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Qian Z, Wang H, Fan H, Lin D, Li W. Three-dimensional digital visualization of phacoemulsification and intraocular lens implantation. Indian J Ophthalmol. 2019;67(3):341–3. https://doi.org/10.4103/ijo.IJO_1012_18.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Suzuki T, Fujishiro T, Sugimoto K, Aihara M. Three-dimensional heads-up surgery in ab-interno trabeculotomy: Image processing-assisted trabeculotomy. PLoS One. 2022 Feb10;17(2):e0263588. doi: https://doi.org/10.1371/journal.pone.0263588.

  13. Bin Helayel H, Al-Mazidi S, AlAkeely A. Can the Three-Dimensional Heads-Up Display Improve Ergonomics, Surgical Performance, and Ophthalmology Training Compared to Conventional Microscopy? Clin Ophthalmol.2021 Feb18;15:679–86. doi: https://doi.org/10.2147/OPTH.S290396.

  14. Zhang Z, Wang L, Wei Y, Fang D, Fan S, Zhang S. The preliminary experiences with three-dimensional Heads-Up Display viewing system for vitreoretinal surgery under various status. Curr Eye Res. 2019;44(1):102–9. https://doi.org/10.1080/02713683.2018.1526305.

    Article  PubMed  Google Scholar 

  15. Kunikata H, Abe T, Nakazawa T. Heads-Up Macular Surgery with a 27-Gauge Microincision Vitrectomy System and Minimal Illumination. Case Rep Ophthalmol. 2016 Nov29;7(3):265–9. doi: https://doi.org/10.1159/000452993.

  16. Nariai Y, Horiguchi M, Mizuguchi T, Sakurai R, Tanikawa A. Comparison of microscopic illumination between a three-dimensional heads-up system and eyepiece in cataract surgery. Eur J Ophthalmol. 2021;31(4):1817–21. https://doi.org/10.1177/1120672120929962.

    Article  PubMed  Google Scholar 

  17. Wang Y, Zhao X, Zhang W, Yang J, Chen Y, THREE-DIMENSIONAL HEAD-UP DISPLAY, VERSUS STANDARD OPERATING MICROSCOPE FOR VITRECTOMY SURGERY. A systematic review and Meta-analysis. Retina. 2022;42(6):1151–60. https://doi.org/10.1097/IAE.0000000000003414.

    Article  PubMed  Google Scholar 

  18. Hoshi S, Tasaki K, Maruo K, Ueno Y, Mori H, Morikawa S et al. Improvement in Dacryoendoscopic Visibility after Image Processing Using Comb-Removal and Image-Sharpening Algorithms. J Clin Med. 2022 Apr 7;11(8):2073. doi: https://doi.org/10.3390/jcm11082073.

  19. Tasaki K, Nishimura T, Hida T, Maruo K, Oshika T. Effects of Image Processing using honeycomb-removal and image-sharpening algorithms on visibility of 27-Gauge endoscopic vitrectomy. J Clin Med. 2022 Sep;26(19):5666. https://doi.org/10.3390/jcm11195666.

  20. Nakajima K, Inoue M, Mizuno M, Koto T, Ishida T, Ozawa H et al. Effects of image-sharpening algorithm on surgical field visibility during 3D heads-up surgery for vitreoretinal diseases. Sci Rep. 2023 Feb16;13(1):2758. doi: https://doi.org/10.1038/s41598-023-29882-5.

  21. Schindelin J, Arganda-Carreras I, Frise E, Kaynig V, Longair M, Pietzsch T, Preibisch S et al. Fiji: an open-source platform for biological-image analysis. Nat Methods. 2012 Jun 28;9(7):676 – 82. doi: https://doi.org/10.1038/nmeth.2019.

  22. Larson MG. Descriptive statistics and graphical displays. Circulation. 2006 Jul 4;114(1):76–81. doi: https://doi.org/10.1161/CIRCULATIONAHA.105.584474.

  23. Selvarasu N, Nachaippan A, Nandhitha NM. Image processing techniques and neural networks for automated cancer analysis from breast thermographs-A review. Indian Journal of Computer Science and Engineering (IJCSE), 2012, 3(1), 133-7. available fromhttp://www.ijcse.com/docs/INDJCSE12-03-01-145.pdf

  24. Wu EX, Cheung MM. MR diffusion kurtosis imaging for neural tissue characterization. NMR Biomed. 2010;23(7):836–48. https://doi.org/10.1002/nbm.1506.

    Article  PubMed  Google Scholar 

  25. Nørhøj Jespersen S. White matter biomarkers from diffusion MRI. J Magn Reson. 2018;291:127–40. https://doi.org/10.1016/j.jmr.2018.03.001.

    Article  CAS  PubMed  Google Scholar 

  26. Park SJ, Do JR, Shin JP, Park DH. Customized Color Settings of digitally assisted vitreoretinal surgery to Enable Use of lower dye concentrations during macular surgery. Front Med (Lausanne). 2022 Jan;24:8:810070. https://doi.org/10.3389/fmed.2021.810070.

  27. Kaur M, Titiyal JS. Three-dimensional heads up display in anterior segment surgeries- expanding frontiers in the COVID-19 era. Indian J Ophthalmol. 2020 Nov;68(11):2338–40. https://doi.org/10.4103/ijo.IJO_2978_20.

  28. Bandoh Y, Qiu G, Okuda M, Daly S, Aach T, Au O. Recent advances in high dynamic range imaging technology. International Conference on Image Processing. 2010, 1109(10), 3125-8. doi: https://doi.org/10.1109/ICIP.2010.5653554.

Download references

Acknowledgements

The authors thank Duco Hamasaki, Professor Emeritus of the Bascom Palmer Eye Institute, University of Miami, Miami, Florida, for the discussions and thorough editing of the manuscript.

Funding

This research received no external funding.

Author information

Authors and Affiliations

Authors

Contributions

K.N., M.I., A.T., Y.Y., M.M., T.K., and T.I. collected data, analyzed the ophthalmological findings. K.N., T.K. and analyzed statistical findings. M.I. and T.O. gave critical suggestions. K.N. and M.I. prepared the manuscript.

Corresponding author

Correspondence to Makoto Inoue.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Institutional Review Committee of the Kyorin University School of Medicine (approval number #1507). The patients received a detailed explanation of the surgical and ophthalmic procedures, and all signed an informed consent form. The patients consented to our review of their medical records and their anonymized use in medical publications. The patient consent for this study was obtained in an opt-out format.

Consent for publication

Not applicable.

Competing interests

K.N. None, M.I.: Research grants from Alcon Laboratories, Inc., Santen Pharmaceutical Co., Ltd., and personal fees (lecture fees) from Alcon Laboratories, Inc., Novartis Pharma K.K., Bayer AG, Carl Zeiss Meditec AG, Novartis Pharma K.K., HOYA Co., Ltd., Chugai Pharmaceutical Co., Ltd., Santen Pharmaceutical Co., Ltd., and Senju Pharmaceutical Co., Ltd., AMO Japan Co., Ltd., Logic and design Co., Ltd., outside the submitted work. A.T. None, Y.Y.: personal fees (lecture fees) from Novartis Pharma K.K., Bayer AG, Carl Zeiss Meditec AG, Novartis Pharma K.K., KOWA, AMO., and CHUGAI PHARMACEUTICAL CO.,LTD, outside the submitted work. M.M.: Personal fees (lecture fees) from Novartis Pharma K.K., Santen Pharmaceutical Co., Ltd., HOYA Co., Ltd., and KOWA Co., Ltd., outside the submitted work., T.K.: Research grants from Ellex, and personal fees (lecture fees) from Alcon Laboratories, Inc., Novartis Pharma K.K., Bayer AG, Carl Zeiss Meditec AG, Novartis Pharma K.K., Santen Pharmaceutical Co., Ltd., HOYA Co., Ltd., Rohto Nitten Co., Ltd., River field Co., Ltd., and Senju Pharmaceutical Co., Ltd., AMO., outside the submitted work., T.I.: Personal fees (lecture fees) from Nikon Co., Ltd., and Santen Pharmaceutical Co., Ltd., outside the submitted work. T.O. received lecture fees and compensation as a consultant from Logic & Design Inc.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nakajima, K., Inoue, M., Takahashi, A. et al. Image sharpening algorithms improve clarity of surgical field during 3D heads-up surgery. Int J Retin Vitr 9, 21 (2023). https://doi.org/10.1186/s40942-023-00462-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40942-023-00462-z

Keywords